. . . a trust for better life care | Log on to: www.anaesthesiaequipment.in , www.lifecareproduct.com

Tuesday, September 28, 2010

LCP Economy Model


LCP-ECON

 AN EXCLUSIVE NITRO LOCKING SYSTEM AT ROTAMETER LEVEL           
INTRODUCTION:-
      LCP’s  Anaesthesia Machine with exclusive Nitrous Oxide locking system has been developed in consultation with number of Anesthetic  & emphasis has been given on the highest level of “PATIENT SAFETY”. Keeping in view the objectives of successfully fulfilling the day to day essential  Anaesthesia needs & requirements of the actual user.
INTEGRATED SAFETY FEATURES :-
      Unlike the limitations of other Nitrous Oxide Lock Systems where presence of supply pressure of oxygen at cylinder level is sufficient to allow flow of Nitrous oxide through its Rotameters.LCP’s  ‘ISF’ works as a true Fail-Safe System as it allows flow of Nitrous Oxide only in presence of oxygen at Rota meter level ,locks the flow of Nitrous Oxide in the absence of flow of oxygen through its Rota meters.
THE APPARATUS :-
      Construction:  Tubular Rigid Steel Section, Powder Coated.
      Self for Ventilator: Sufficient space provided at the top level of lower drawer to place ventilator.
      Table Top: Guarded Table Top Provided with a stainless Tray.
       Instrument Tray: Top Tray  Provided at eye level for keeping Monitoring Equipments.

ATTACHMENT:-
     ROTAMETERS: 230 mm long accurately calibrated rotating bobbin flow meters.
        Oxygen:            100 ml/min to 8 LPM
        Nitrous Oxide:  200 ml/min to 12 LPM
       EATHER VAPORIZER: The Vaporizer of Eather or Halothane is supplied as standard accessories.
       EMERGENCY UNIT: Emergency oxygen flush button is provided at the back bar & guarded to avoid accidental passing.    FLOW RANGE: 35LPM calibrated at 60 PSI (+5%,-5%)

PATIENT CIRCUIT:-  Standard Magill’s Circuit consisting of Bag Mount – 01 No corrugated Hose -01 No.Angle Mount -01 No., Catheter Mount -01 No., Face Mask -01 No.,Rebreathing Bag -01 No.,Key Spanner -02 No.

OPTIONAL:-
      Circle Absorber System:  Twin Canister Circle Absorber System is provided with Swiveling facility. Canisters & inspiratory valve have quick – fit Tapper fitting with Rubber  ‘O’  ring for leak proof seal & easy maneuver ability. Bag Mount & APL valve Assembly have swivel arms for easy accessibility.